Sampling methods: | |
Pollutants: | |
Others: | |
Top searches: |
It applies pitot tube isokinetic sampling method to collect particulate matter in stationary source emission, applies filtering and weighting method to determine the quality and applies constant potential electrolysis method to determine the concentrations of harmful gases. It can be used to determine the stack dust (gas) emissions concentration, equivalent concentration, total amount and dust removal, desulfurization efficiency of variety of boilers and furnaces. It can also be used to measure stack dynamic pressure, static pressure, the pressure and temperature in front of flow meter, temperature of flue gas, humidity ratio, concentrations of O2, SO2, CO, NO, NO2, H2S, CO2, Isokinetic sucking velocity and so on.
It is widely used for environmental protection, testing company, Industrial And Mining Enterprises, hygiene, working place, safety inspection, military, scientific research, education and other departments.
HJ/T 48-1999 Technical conditions of sampler for stack dust
JJG 968-2002 Verification regulation of flue gas analyzers
JJG 680-2007 Verification regulation of samplers for stack dust
DB37/T 2537-2014 Determination of low concentration particulate matters in stationary source emission of Shandong province gravimetric method
When the concentration of particulate matters in stationary source emission is lower than 50mg/m3 during low concentration determination, as the sampling volume under standard condition is 1m3,the lower testing limit is 1mg/m3
High-load, low-noise, high flow-rate sampling pump, up to 100L/min
Isokinetic tracking sampling controlled by microcomputer, with exclusive way of adjustment and short response time
Accurate electronic flow meter control, real-time temperature and pressure monitoring, automatic flow rate adjustment
Wide temperature large-screen reversible TFT color display, with resistance to high cold and wide visual angle,accomplished man-machine interactive interface
Precise design of dust-proof and water-proof industrial board, suitable for severe environment, easy to operate.
Unique design of efficient gas-water separator with high efficiency of drying, improves silicone's use
Design of Visible dust-filter core, with convenience for replacement and efficiency to filter the particulate matters, offering further protection of gas circuit and sampling pump
Automatic fault detection function and airtightness alarm function, easy for user's maintenance and usage
Industrial high-speed imbedded IPC core, Run WINCE operating system
Intelligent software calibrating function
Developed Windows environment based computer database and communication system, accomplished storage, query and print-out through computer communication.
Real-time display the curve of the sampled data, easy to observe the changes in the sampling data;
With DC power interface, accomplished AC/DC power supply
Automatic power monitoring function: If power off during sampling, it can automatically save data and continue to sample after power resume
Isolated DIO channel and sampling probe earth line design, with strong ability of anti-static and anti-interference
Positive sampling nozzle and negative sampling nozzle of pitot tube are connected to the mainframe with different colored channels to prevent reversal, accurately determine dynamic pressure
Automatically Prevent the dust in filter cartridge from being sucked back after completion of sampling
Super high-capacity data storage, up to more than 100000 groups
Various man-machine interface: Supports mouse, U-disk, keyboard, touchpad, and printer and so on.
Adopt high speed miniature thermal printer, with high-speed, low-noise and high-accuracy
Support Chinese input, automatically memorize information of stack condition configuration, convenient for later use.
small size, light weight, easy to operate, easy to carry
Flue gas processor is optional, to increase the accuracy of flue gas composition test
Low concentration stack dust multi-function sampling probe is optional, to accomplish low concentration PM sampling
Pitch smoke sampling is optional, to measure the emission concentration of pitch smoke.
Oil smoke sampling probe is optional, to measure the emission concentration of oil smoke for catering industry
The above parameters are totally up to the requirements of the relevant national standards. If product upgrades, or image does not match with real machine, please refer to the real machine.
If you have special needs, please call:
the unified national customer sales hotline:400-676-5892 or service hotline: 400-676-6892 to consulate details, Thank you for your cooperation!